EasyManua.ls Logo

THERA-Trainer BEMO - Installation and movement (transport)

THERA-Trainer BEMO
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN User manual THERA-Trainer bemo
23
Article no. A007-661_Rev 2.0 - Last edited: 2020-06-02
12. Installation and movement (transport)
CAUTION!
Risk of injury due to incorrect transport!
Only transport the THERA-Trainer with double steering castors on level and firm ground.
When installing the THERA-Trainer, make sure that the trainer is placed on a level, firm and non-slip surface
and that the parking brakes are locked. Reposition the THERA-Trainer if the position of the trainer changes
during therapy.
Stop training session.
Reposition the THERA-Trainer (see chapter 17.4).
Continue training.
To transport the THERA-Trainer:
Adjust height to be able to move the trainer over the bed/couch.
Disconnect power plug from the mains supply.
Hang the mains cable on the THERA-Trainer hook.
Disconnect the equipotential bonding cable from the THERA-Trainer.
Release parking brakes.
Transport the THERA-Trainer to the next place of use.
Hold the THERA-Trainer by the lifting column, crossbeam or drive unit to guide the trainer during
transport.
Make sure that the THERA-Trainer fits through doors/corridors.
The manufacturer recommends removing options such as foot rests, therapy grips, arm rests,
etc. for transport.
13. Maintenance/repairs/reuse
WARNING!
Danger to life due to electric shock!
Prior to any maintenance or repair work, disconnect the power plug.
WARNING!
Danger to life due to improper maintenance!
Service or repair work should only be carried out by trained suppliers.
The THERA-Trainer is maintenance-free.
In accordance with Section 11 of the German Medical Devices Operator Ordinance (MPBetreibV), no safety
check is required. The manufacturer recommends performing a safety check every 24 months. If the housing
of the THERA-Trainer is opened or work is carried out on the electronics, a safety check is necessary.
The relevant information for the safety check is available from the manufacturer upon request.
The THERA-Trainer is suitable for re-use by a different patient. Before the THERA-Trainer is reused:
Clean and disinfect the device thoroughly (see chapter 19.3).
Perform a visual check or a function check.
Check if all accessories and options are complete.

Table of Contents