EN User manual THERA-Trainer bemo
31
Article no. A007-661_Rev 2.0 - Last edited: 2020-06-02
Training is only possible if a patient emergency stop or dummy plug for the patient emergency stop is
connected. Before training, check whether the patient emergency stop or dummy plug for the patient
emergency stop is connected and working. Otherwise the trainer will display error E556 / E557.
17.1. Switching on
CAUTION!
Risk of injury due to the THERA-Trainer starting up!
Make sure that no arms or legs are in the THERA-Trainer during start-up.
Connect power plug to socket.
o Only use the power cable supplied by the manufacturer.
Operating system boots up.
Display while the operating system boots (automatically one after the other):
1. All elements of the control and display unit that can be displayed (for function control)
2. Current setting of the sub menu
3. Software version of the control and display unit
4. Software version of the main controller board
5. Operating hours
WARNING!
Risk of injury due to defective T.assist!
Only train under supervision if T.assist is detected as defective during boot-up:
Make sure that status indicators light up in the colours blue, red and green every time you start up the
device.
Make sure that beeps sound every time you start up the device.
Make sure that no error messages are displayed every time you start up the device.
Have malfunctions repaired by trained suppliers.
Check T.assist during each boot-up.
After the operating system has booted up, the THERA-Trainer is ready for operation.
17.2. Stand-by mode
The THERA-Trainer switches to stand-by mode when no button is pressed for approximately 5 minutes
(not during training).
A beep will sound and the screen backlight will turn off.
To activate the THERA-Trainer:
Press any button on the control and display unit.
Operating system boots up. After the operating system has booted up, the THERA-Trainer is ready for
operation.
CAUTION!
Risk of injury from unsuitable chair/wheelchair while exercising upper body in a seated position!
Make sure that the chair/wheelchair does not tip over or slip.
Only use chairs/wheelchairs with backrests of sufficient height.

Table of Contents