24
Article no. A007-661_Rev 2.0 - Last edited: 2020-06-02
14. Notes on electromagnetic compatibility
14.1. Cables, cable lengths and accessories
Only operate the THERA-Trainer with the original mains supply cable.
14.2. Warning notice regarding installation
Do not use the THERA-Trainer in the immediate vicinity of other devices or stacked with other devices.
If there is no alternative to operation next to or stacked with other devices, observe the THERA-Trainer
to check if the arrangement permits it to work as intended.
14.3. Electromagnetic emission and immunity
The device is classified in Group 1 Class B in accordance with CISPR 11 and designed for interference
immunity testing levels in accordance with EN 60601-2:2016-05 (EN 6060 1-2:2015, IEC 60601-1-2:2014)
for domestic healthcare environments described in Tables 4, 5, 8 and 9.
Deviations or simplifications regarding the above-mentioned immunity test levels are not permitted.
In order to maintain basic safety and the essential performance characteristics within the expected service life
of 10 years, no precautions (maintenance activities) are required with regard to electromagnetic disturbances.
Table 4 - Casing:
Phenomenon
Basic standard
Interference immunity test levels
Domestic healthcare environment
Discharge of static electricity
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV Air
High-frequency
electromagnetic fields
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
High-frequency
electromagnetic fields in the
immediate vicinity of wireless
communication devices
IEC 61000-4-3
In accordance with DIN EN 60601-1-2:2016,
EN 60601-2:2016-05 (EN 6060 1-2:2015, IEC 60601-1-2:2014)
for healthcare environments described in Table 9
Magnetic fields with power
rated frequencies
IEC 61000-4-8
30 A/m
50 Hz