60
TROUBLESHOOTING
First Octet is 1-223 Range Displayed when the first octet of IP address, default gateway, primary DNS
server or secondary DNS server is set to a value out of the specified input
range. Enter a value within the input range.
Value is irregular.
Input valid value
Displayed when the "Subnet mask" of the "LAN connection" is set to a value off
the input rule. Enter a value within the "Subnet mask" input rule.
The IP address is 0-255 Range Displayed when any one of the respective octets is set to a value out of the
specified input range. Enter a value within the input range.
Prefix of exam. ID need 3 letter Displayed when the input examiner ID prefix is less than 3 characters. Enter a
prefix with 3 characters.
Not correct password Displayed when the password inputted to select a special mode is incorrect.
Please wait until packing mode is
finished...
Indicates that the packing operation is in process. Wait until it is completed.
Please check the DATE/TIME Displayed when the battery for the built-in clock becomes run down.
- When the battery consumed, confirm the difference in time and adjust it.
- When the battery becomes completely drained, to verify whether time
stopping occurred and call your service engineer.
Cannot detect x position.
Please turn the switch off/on.
Displayed when the auto alignment sensor and the machine are not correctly
connected or not connected at all.
When this message is displayed repeatedly, even if you turn the switch on
again, please call your service engineer.
Cannot detect y position.
Please turn the switch off/on.
Displayed when the auto alignment sensor and the machine are not correctly
connected or not connected at all.
When this message is displayed repeatedly, even if you turn the switch on
again, please call your service engineer.
Cannot detect z position.
Please turn the switch off/on.
Displayed when the auto alignment sensor and the machine are not correctly
connected or not connected at all.
When this message is displayed repeatedly, even if you turn the switch on
again, please call your service engineer.
Failed to initialize TF motor.
Please turn the switch off/on.
Displayed when the fixation target sensor and the machine are not correctly
connected or not connected at all.
When this message is displayed repeatedly, even if you turn the switch on
again, please call your service engineer.