- 13 - MODEL 3587
3.7 Memory setting
Input of MEMORY SET
① Press F1 [SET] key in standby state.
Display changes to SETTING MODE.
② Press F1 [SEL] key to select MEMORY SET.
Memory setting can be changed by pressing F4 [ENT] key.
Basic operation
Memory No., measurement mode, measurement voltage, resistance range, comparator,
timer, mask timer can be set.
Operation key
F1 [SEL] : Selection of setting item.
The back ground of the display color of the selected item will be
changed.
F2 [▲] : Change of setting values
F3 [▼] : Selection of setting from TEST MODE ,RANGE ,COMP RNG
setting value can be increased or decreased by numerical setting.
F4 [ENT] :Memorize the setting and the display returns to measurement
standby state.
HIGH
GOOD
LOW
START STOP
3587MODEL
LOW
REMOT E
GUARD
HIGH
[]
F3F2F1
SETTING MODE
SYSTEM SET
EXT. POLARITY : NPN
[ENT]
1
2
MEMORY SET
TEST
[RCL][SEL]
F4
GENERAL SET
HIGH
GOOD
LOW
START STOP
3587MODEL
LOW
REMOTE
GUARD
HIGH
F3F2F1
[ ENT ]
MEMORY SET
MEMORY No
TEST M ODE
VOLTAG E
RANGE
AUTO
1000V
2000MΩ
1/3
[][]
1
TEST
[SEL]
F4