MODEL 3587 - 39 -
6.2 List of communication command and response
Comman
Response Contents
DATA? □
Cr
□
Lf
DATA=00.12MOHM,HIGH,R
□
Cr
□
Lf
① ② ③ ④ ⑤
Read measurement data
①Response ②Measured value ③Judgment result ④Operation condition ⑤Delimiter
DATA=9990 MOHM,HIGH,R□
Cr
□
Lf
DATA=0.000MOHM,LOW ,R
□
Cr
□
Lf
DATA=12.34MOHM,GOOD,R
□
Cr
□
Lf
DATA=OVERMOHM,NULL,R
□
Cr
□
Lf
DATA=123.4MOHM,NULL,T
□
Cr
□
Lf
DATA=12.34MOHM,GOOD,T
□
Cr
□
Lf
Measured value 9990MΩ HIGH Judgment Stand-by (Measurement end)
Measured value 0.000MΩ LOW Judgment Stand-by (Measurement end)
Measured value 12.34MΩ GOOD Judgment Stand-by (Measurement end)
Measured value OVER Judgment off Stand-by (Measurement end)
・・・Judgment OFF by STOP
Measured value 123.4MΩ Judgment off Under measurement
AUTO mode
Measured value 12.34MΩ GOOD Judgment Under measurement
・・・CONTINUE mode
TEST? □
Cr
□
Lf
TEST=TEST
□
Cr
□
Lf
① ② ③
Read operation condition
①Response ②Operation condition ③Delimiter
TEST=READY□
Cr
□
Lf
TEST=TEST
□
Cr
□
Lf
Stand-by condition (Measurement end)
Under measurement