MODEL 3587 - 32 -
5.1.2 Connection
Compatible electrical wire: AWG26 to AWG20
Side of electrical wire is peeled off as shown in figure below and wire is inserted into the
terminal by pressing the release button down with help of screwdriver.
Peel length8 to 9mm
5.1.3 PNP/NPN switching
Input / output signal of PNP (source) and NPN (sink) located at rear side of tester can be
changed according to the interface to be connected.
Set the switch to P side during PNP.
Set the switch to N side during NPN
5.1.4 Input/output signal
Input signal
Input ON residual voltage Less than 1V ON current less than 30 mA
Response time 5 ms (MAX)
Output signal
Signal Open drain output (nonpolar)
Max. load DC30V 30mA
Residual voltage Less than 1V During load current: 30mA
8~9mm
100-240V 50-60Hz
AC LINE 28VA MAX
O- COM
GO
LO
HI
CH G
C- ERR
TE ST
I- COM
ME M1
ME M2
ME M4
ME M8
ST ART
RE SET
CONTROL
PN
RS-232C
LOW
HIGH
L CONTACT
H CONTACT
BCDOCOUT
P/N selector switch