- 31 - MODEL 3587
5. External control
Output signal of start / stop selection of memory, judgment and TEST etc. Can be
obtained from the control terminal located at rear side of the tester.
5.1 Control terminal
5.1.1 Terminal arrangement
No.
Signal
name
Input/
output
Function
1 O-COM ―
Output common
2 to 7 Commo
It is internally connected with I-COM No.8.
2 GO Output
Comparator
It is output of GOOD judgment. Output of ON is obtained
during GOOD judgment.
3 LO Output
Comparator
It is output of LOW judgment. Output of ON is obtained
during LOW judgment.
4 HI Output
Comparator
It is output of HIGH judgment. Output of ON is obtained
during HIGH judgment.
5 CHG Output
ON output is obtained when the voltage of the HIGH
measuring terminal become more than 10V.
6 C-ERR Output
It is connection check measurement result output signal.
ON output is obtained during NG result of connection check.
7 TEST Output
It is output signal during measurement operation.
During test voltage output, ON output of connection check is
obtained.
8 I-COM ―
Input common
9 to 14 Common
It is internally connected with O-COM No. 1.
9
10
11
12
MEM1
MEM2
MEM4
MEM8
Input
Memory is called by inputting memory No.
Refer to the table of memory operation (Section 5.1.5) for
selection of memory.
13 START
Input
Measurement start signal
Measurement starts when input is ON during standby state.
START signal operates if start input is set to REMOTE1.
14 RESET
Input
Stop measurement, return judgment signal
If ON during measurement, measurement stops.
When input is ON during standby state, reset (OFF)
judgment is obtained.