- 15 - MODEL 3587
3.7.3 Measurement voltage setting
Press F1 [SEL] to select MEMEORY SET and press F4 [ENT] to enter it.
Select VOLTAGE by F1 [SEL] key.
Press F2 [▲], F3 [▼] key.to change the setting
If keys are pressed continuously, the speed of setting value changes in 3 steps.
Setting range 25V to 1050V
Press F4 [ENT] to memorize the setting and the display returns to measurement
standby state.
When the measurement range is out of range, OUT OF RANGE is displayed for about
1 s and display returns to setting mode.
Set the measurement voltage or measurement range within the range of Table 3.1.
To continue the other setting, select the items that you want to change by pressing
F1 [SEL].
Table 3.1 Voltage setting and resistance range
Measurement
voltage setting
Resistance range
25V to 99V 2MΩ, 20MΩ, 200MΩ, AUTO
100V to 499V 2MΩ, 20MΩ, 200MΩ, 2000MΩ, AUTO
500V to 1050V 20MΩ, 200MΩ, 2000MΩ, AUTO
HIGH
GOOD
LOW
START STOP
MODE
LO
REMOTE
GUARD
IGH
F3F2F1
[ENT]
MEMORY SET
MEMORY No
TEST MODE
VOLTAGE
RANGE
1
AUTO
2000MΩ
1/3
[][]
1000V
TEST
[ SEL ]
F4