- 19 - MODEL 3587
3.7.7 Timer/mask timer
Measurement time is set in timer when the measurement time is carried out in AUTO
mode.
Mask timer is set when the measurement start and judgment are not carried out by AUTO
mode.
Setting range
Timer is set to 00.2s to 99.9 s
The mask timer is set to 00.1s to 99.9 s /OFF
Mask timer can’t be operated if mask time is set to OFF.
Setting method
Press F1 [SEL] key to select MEMORY SET and press F4 [ENT] key to enter it.
Select TIMER or MASKTIMSR by F1 [SEL] key.
Press F2 [▲], F3 [▼] key to change the setting
If the key is pressed continuously, speed can be changed in 3 stages.
Setting range TIMER 00.2 to 99.9 s
MASK TIMER 00.2 to 99.9 s /OFF
Press F4 [ENT] to memorize the setting by displays returns to measurement mode.
When MASKTIME > TIMER, OUT OF RANGE is displayed for approximately 1 s
and display returns to setting mode.
MASKTIME TIMER, Set the values.
To continue the other setting, select the items that you want to change by pressing F1
[SEL].
HIGH
GOOD
LOW
START STOP
MODE
LOW
REMOTE
GUARD
IGH
F3F2F1
[ENT]
MEMORY SET
3/3
TIMER
MASKTIMER 00.5s
[][]
10.0s