MODEL 3587 - 28 -
4.2 REMOTE / MANUAL
4.2.1 MANUAL operation (MANUAL)
In manual operation, the start and finish of the measurement are done by the START and
STOP key located on the front panel.
Press START key to start the measurement.
As the measurement starts, the measuring voltage is applied to the measurement
terminal and the TEST LAMP is lit up.
To finish the measurement, press STOP key or ON the RESET terminal.
In case that the measurement mode is AUTO operation, the measurement
automatically finishes with the time up of the timer or NG judgement.
4.2.2 Remote 1 operation (REMOTE 1)
During REMOTE 1 operation, the start and finish of the measurement are done by the
START and RESET terminals located on the rear side of the tester.
To start the measurement, ON the START (minimum width 5ms).
As the measurement starts. The measuring voltage is applied to the measuring
terminal, and the TEST lamp is lit up.
To finish the measurement, ON the RESET terminal (minimum width 5ms), or press
the STOP key.
In case that the measurement mode is AUTO operation, the measurement
automatically finishes with the time up of the timer or NG judgement.
4.2.3 Remote 2 operation (REMOTE 2)
During REMOTE 2 operation, the start and finish of the measurement are done by ON
and OFF of the probe switch (option) respectively.
To start the measurement, press the probe switch continuously.
As the measurement starts, measurement voltage is applied to the measurement
terminals and TEST LAMP is lit up.
To finish the measurement, release the probe switch.
In case that the measurement mode is AUTO operation, the measurement
automatically finishes with the time up of the timer or NG judgement.