MODEL 3587 - 30 -
4.4 Automatic discharge function
If the specimen possesses capacitive components, it may be charged with the test voltage
and remaining as residual voltage may cause an electric shock accident.
A discharging circuit (discharging resistance approximately 100KΩ) is built-in in this
tester to discharge the residual voltage.
Measurement is completed without removing the connection with specimen.
With the specimen being connected, the discharging circuit built-in tester discharges
the residual voltage specimen
During the discharging, TEST lamp blinks, and it turns off at voltage less than about
10V.
The discharging time becomes longer depending upon the volume of capacitive
components.
Automatic discharge function can be turned ON / OFF by setting on automatic discharge
function (Section 3.9.1) on system setting.