EasyManua.ls Logo

Tsuruga 3587 - RS-232 C Operation and Commands

Default Icon
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODEL 3587 - 42 -
Comman
d
Response Contents
VOLT=1000V
Cr
Lf
VOLT= 25V
Cr
Lf
VOLT= 100V
Cr
Lf
VOLT= 500V
Cr
Lf
VOLT=1000V
Cr
Lf
VOLT= 20V
Cr
Lf
VOLT=1000V
Cr
Lf
Set the measurement voltage at memory No. being displayed
Response Set voltage Delimiter
VOLT= 25V
Cr
Lf
VOLT= 100V
Cr
Lf
VOLT= 500V
Cr
Lf
VOLT=1000V
Cr
Lf
VOLT=ERR
Cr
Lf
Set voltage DC25V (Set voltage range 25V to 1000V, Step 1V)
Set voltage DC100V
Set voltage DC500V
Set voltage DC1000V
Out of voltage setting range error
During setting of voltage 25 to 99V at measurement range of 2000MΩ, range changes to
200MΩ
During setting of voltage 500 to 1000V at measurement range 2MΩ, range changes to
20MΩ
TIMER?
Cr
Lf
TIMER=10.0
Cr
Lf
Read the setting of timer of memory No. being displayed
Response Setting of timer Delimiter
TIMER=00.2
Cr
Lf
TIMER=01.0
Cr
Lf
TIMER=99.9
Cr
Lf
Set timer 0.2 s
Set timer 1.0 s
Set timer 99.9 s
TIMER=10.0
Cr
Lf
TIMER=00.2
Cr
Lf
TIMER=01.0
Cr
Lf
TIMER=99.9
Cr
Lf
TIMER=0.00
Cr
Lf
TIMER=10.0
Cr
Lf
Set the timer of memory No. being displayed.
Response Setting of timer Delimiter
TIMER=00.2
Cr
Lf
TIMER=01.0
Cr
Lf
TIMER=99.9
Cr
Lf
TIMER=ERR
Cr
Lf
Set timer 0.2 s (Setting range 0.02 to 99.9 s, Step 0.1s)
Set timer 1.0 s
Set timer 99.9 s
Out of timer setting range error