MODEL 3587 - 42 -
Comman
Response Contents
VOLT=1000V□
Cr
□
Lf
VOLT= 25V
□
Cr
□
Lf
VOLT= 100V
□
Cr
□
Lf
VOLT= 500V
□
Cr
□
Lf
VOLT=1000V
□
Cr
□
Lf
VOLT= 20V
□
Cr
□
Lf
VOLT=1000V
□
Cr
□
Lf
① ② ③
Set the measurement voltage at memory No. being displayed
①Response ②Set voltage ③Delimiter
VOLT= 25V□
Cr
□
Lf
VOLT= 100V
□
Cr
□
Lf
VOLT= 500V
□
Cr
□
Lf
VOLT=1000V
□
Cr
□
Lf
VOLT=ERR
□
Cr
□
Lf
Set voltage DC25V (Set voltage range 25V to 1000V, Step 1V)
Set voltage DC100V
Set voltage DC500V
Set voltage DC1000V
Out of voltage setting range error
During setting of voltage 25 to 99V at measurement range of 2000MΩ, range changes to
200MΩ
During setting of voltage 500 to 1000V at measurement range 2MΩ, range changes to
20MΩ
TIMER? □
Cr
□
Lf
TIMER=10.0□
Cr
□
Lf
① ② ③
Read the setting of timer of memory No. being displayed
①Response ②Setting of timer ③Delimiter
TIMER=00.2□
Cr
□
Lf
TIMER=01.0
□
Cr
□
Lf
TIMER=99.9
□
Cr
□
Lf
Set timer 0.2 s
Set timer 1.0 s
Set timer 99.9 s
TIMER=10.0□
Cr
□
Lf
TIMER=00.2
□
Cr
□
Lf
TIMER=01.0
□
Cr
□
Lf
TIMER=99.9
□
Cr
□
Lf
TIMER=0.00
□
Cr
□
Lf
TIMER=10.0
□
Cr
□
Lf
① ② ③
Set the timer of memory No. being displayed.
①Response ②Setting of timer ③Delimiter
TIMER=00.2□
Cr
□
Lf
TIMER=01.0
□
Cr
□
Lf
TIMER=99.9
□
Cr
□
Lf
TIMER=ERR
□
Cr
□
Lf
Set timer 0.2 s (Setting range 0.02 to 99.9 s, Step 0.1s)
Set timer 1.0 s
Set timer 99.9 s
Out of timer setting range error