EasyManua.ls Logo

Tsuruga 3587 - Page 48

Default Icon
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODEL 3587 - 44 -
Comman
d
Response Contents
MEM? MEM=01
Cr
Lf
Read memory No. being displayed
Response Memory No. Delimiter
MEM=01
Cr
Lf
MEM=10
Cr
Lf
Memory No.01
Memory No.10
MEM=CALL01
Cr
Lf
MEM=CALL01
Cr
Lf
Call the specified set No.
Response Memory No. Delimiter
MEM=CALL01
Cr
Lf
MEM=CALL10
Cr
Lf
MEM=CALL29
Cr
Lf
Memory No.01 (Setting range 01 to 10)
Memory No.10
Error when out of setting range
WRITEMEMORY
Cr
Lf
Write set memory 01 to 10 to EEPROM
WRITE SUCCESS
Cr
Lf
WRITE ERR
Cr
Lf
Successful EEPROM writing
Failure of EEPROM writing
START
Cr
Lf
START
Cr
Lf
START ERR
Cr
Lf
Start the measurement
ERR is obtained during REMOTE/MANUAL key setting to REMOTE (No
measurement start)
STOP
Cr
Lf
STOP
Cr
Lf
Undergoing measurement cancellation
During stand-by state, judgment result becomes clear (Make OFF)
ONLINE?
Cr
Lf
ONLINE=ON
Cr
Lf
Read online status
Response State Delimiter
ONLINE=ON
Cr
Lf
ONLINE=OFF
Cr
Lf
Online is ON
Online is OFF
ONLINE=ON
Cr
Lf
ONLINE=ON
Cr
Lf
ONLINE=OFF
Cr
Lf
ONLINE=XYZ
Cr
Lf
ONLINE=ON
Cr
Lf
Set online
Response State Delimiter
ONLINE=ON
Cr
Lf
ONLINE=OFF
Cr
Lf
ONLINE=ERR
Cr
Lf
Set online to ON
Set online to OFF
Setting error of online