Ulvac C30HVRT - Figure 2-10 Timing Chart of C30 HVRT When a Short Power Interruption Occurs

Default Icon
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 2. Considerations before Installation
2-12
ULVAC CRYOGENICS INCORPORATED
INPUT POWER SUPPLY
COMPCRYOCOOLER
START
COMPCRYOCOOLER
STOP
COMPCRYOCOOLER
ACK
COMP OPERATION
CRYOCOOLER OPERATION
Figure 2-10 Timing chart of C30HVRT when a short power interruption occurs
(In Momentary operating mode)
(Operation after a short power interruption)
While the compressor is running in “Momentary operating mode”, if a short power
interruption occurs and the input power is recovered within two seconds, the
compressor pump restarts the operation automatically.
For C30HVRT, cryocooler operation is synchronized with the compressor operation as
shown in Figure 2-10.
In “Alternate operating mode”, the compressor restarts as soon as the input power is
recovered because there is always the command from the equipment side.
NOTE: Refer to “Section 4.5.2 Connecting remote cables” for remote signal wiring.