User's Guide _______________________________________________________________________
88
______________________________________________________ M210296en-A
the command initializes the two-point calibration sequence, where two
temperatures must be simulated.
When you type
TCAL TS DTS14B_temperature
a single-point calibration to the TS is made. That is, the scaling factor
TS 0 is adjusted by the command routine.
The following command
TCAL TS 0.0
makes a zero calibration, if the temperature sensor DTS14B is in an
ice bath or otherwise at a temperature of 0 °C .
The following command
TCAL TS 0.0581 -59.0
sets both scaling factors.
The system output is as follows:
DRI TEMPERATURE SCALES
TS 1 0.0581 TS 0 -59.0000 TS 2.8
CLEAN Command
The CLEAN command has no parameters and it is used to set the
clean references for contamination control. This command is given
during maintenance procedures after cleaning the lenses or after
replacing the transmitter or receiver board.
When you type
CLEAN
the FD12P output is as follows:
CLEAN REFERENCES
TRANSMITTER 12.0
RECEIVER 1402
UPDATED