Using the RS-232 Serial Port
Setup, Tests, and Adjustments
Force FX-8C Service Manual 5-41
When activation ceases, the following message is displayed:
key request removed
Step 2 – Enter the Commands
You can enter commands to change settings or display data about the
generator. Each command should begin with a colon (:) and end with
Enter. The # symbol represents a numeric value that you enter with
specific commands as noted.
Data Displays
Some commands result in the display of data for each mode, including
CEM. The display format is as follows:
In data displays, numbers represent the Force FX-8C modes. You also use
these numbers when you enter commands to select a specific mode.
Bipolar Modes
0 Precise
1 Standard
2 Macro
Cut Modes
0 Low
1 Pure
2 Blend
Coag Modes
0 Desiccate
1 Fulgurate
2 Spray
Bipolar Precise Bipolar Standard Bipolar Macro
Default
Fulgurate
Cut Low Cut Pure Cut Blend Cut CEM
Coag Desiccate
LCF Fulgurate
Coag Spray Coag CEM