Performance Characteristics
3-6 Force FX-8C Service Manual
Low Frequency (50–60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Enclosure source current,
ground open
< 300 µA
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: < 50 µA
Sink current at high line,
all inputs
< 50 µA
Bipolar RF leakage current < 59.2 mA
rms
Monopolar RF leakage
current
(additional tolerance)
< 150 mA
rms
CEM output modes < 150 mA
rms
at ≤ 50 W