Standards and IEC Classifications
3-8 Force FX-8C Service Manual
Standards and IEC Classifications
Class I Equipment (IEC 601-1)
Accessible conductive parts cannot become live in the event of a basic
insulation failure because of the way in which they are connected to the
protective earth conductor.
Type CF Equipment (IEC 601-1)/Defibrillator Proof
ATTENTION
Consult accompanying documents.
The generator output is floating (isolated) with respect to
ground.
DANGER
Explosion risk if used with flammable anesthetics.
To reduce the risk of electric shock, do not remove the cover.
Refer servicing to qualified service personnel.
The Force FX-8C generator provides a high degree of protection
against electric shock, particularly regarding allowable leakage
currents. It is type CF isolated (floating) output and may be
used for procedures involving the heart.
The Force FX-8C generator patient return electrode terminal is
protected from defibrillator discharge according to
ANSI/AAMI HF18 and IEC 601-2-2.
F