Output Characteristics
3-12 Force FX-8C Service Manual
Monopolar Coag
CEM Cut
CEM Coag
Output Waveforms
Instant Response Technology, an automatic adjustment, controls all
bipolar modes and all cut modes. It does not control the coag modes
because of their fulguration capabilities. As tissue resistance increases
from zero, the generator outputs constant current followed by constant
power followed by constant voltage. The maximum output voltage is
controlled to reduce capacitive coupling and video interference and to
minimize sparking.
12345678910
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
31 32 33 34 35 36 37 38 39 40
45 50 55 60 65 70 75 80 85 90
95 100 110 120
12345678910
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
31 32 33 34 35 36 37 38 39 40
45 50 55 60 65 70 75 80 85 90
95 100
12345678910
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
31 32 33 34 35 36 37 38 39 40
45 50 55 60 65 70