EasyManua.ls Logo

Valleylab Force FX-8C - Input Power

Valleylab Force FX-8C
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Characteristics
Technical Specifications
Force FX-8C Service Manual 3-7
Input Power
110–120 Volt 220–240 Volt
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Input mains voltage, full regulation range:
104–132 Vac
Input mains voltage, full regulation range:
208–264 Vac
Input mains voltage, operating range:
85–132 Vac
Input mains voltage, operating range:
170–264 Vac
Mains current (maximum):
Idle: 0.4 A
Bipolar: 2.0 A
Cut: 7.0 A
Coag: 4.0 A
Mains current (maximum):
Idle: 0.2 A
Bipolar: 1.0 A
Cut: 3.5 A
Coag: 2.0 A
Mains line frequency range (nominal):
50 to 60 Hz
Mains line frequency range (nominal):
50 to 60 Hz
Fuses (2): F8 A Fuses (2): T4 A
Power cord: 3-prong hospital grade
connector
Power cord: 3-prong locally approved
connector

Table of Contents

Related product manuals