Performance Characteristics
3-2 Force FX-8C Service Manual
Dimensions and Weight
Operating Parameters
Transport and Storage
Width: 35.6 cm (14 in.)
Depth 45.7 cm (18 in.)
Height 11.1 cm (4 3/8 in.)
Weight < 8.2 kg (< 18 lbs.)
Ambient temperature
range
10° to 40° C (50° to 104° F)
Relative humidity 30% to 75%, noncondensing
Atmospheric pressure 700 to 1060 millibars
Warm-up time If transported or stored at temperatures outside the
operating temperature range, allow one hour for the
generator to reach room temperature before use.
Ambient temperature
range
-40° to 70° C (-40° to 158° F)
Relative humidity 10% to 100%, condensing
Atmospheric pressure 500 to 1060 millibars
Duration of storage If stored longer than one year, the battery must be replaced
and a full checkout, including calibration, must be
completed before use. For instructions, refer to Section 5,
Calibrating the Generator, in this manual.