Setting Up the Generator
5-2 Force FX-8C Service Manual
Setting Up the Generator
1. Verify the generator is off by pressing the power switch off (O).
2. Place the generator on a stable flat surface, such as a table, platform, or
Valleylab cart. Carts with conductive wheels are recommended. For
details, refer to the procedures for your institution or to local codes.
Provide at least four to six inches of space from the sides and top of the
generator for cooling. Normally, the top, sides, and rear panel are
warm when the generator is used continuously for extended periods
of time.
Ensure that the generator rests securely on the cart or platform. The
underside of the generator contains four rubber feet and additional
holes that allow you to reposition the feet to ensure stability. Use a
Phillips screwdriver to remove the rubber feet from the generator.
Then, reinstall the feet in the preferred location.
3. According to the procedures used by your institution, connect an
equipotential grounding cable to the grounding lug on the rear panel
of the generator. Then, connect the cable to earth ground.
4. Plug the generator power cord into the rear panel receptacle.
5. Plug the generator power cord into a grounded receptacle.
6. Turn on the generator by pressing the power switch on (|). Verify the
following:
• All visual indicators and displays on the front panel illuminate.
Warning
Electric Shock Hazard Connect the generator power cord to a properly grounded
receptacle. Do not use power plug adapters.
Fire Hazard Do not use extension cords.
Caution
Do not stack equipment on top of the generator or place the generator on top of
electrical equipment (except a Force GSU unit, a Force Argon unit, a CUSA
System 200, or a CUSA EXcel unit). These configurations are unstable and/or do
not allow for adequate cooling.
Provide as much distance as possible between the electrosurgical generator and
other electronic equipment (such as monitors). An activated electrosurgical
generator may cause interference with them.
Notice
If required by local codes, connect the generator to the hospital equalization
connector with an equipotential cable.
Connect the power cord to a wall outlet having the correct voltage. Otherwise
product damage may result.