Setting Up the Generator
Setup, Tests, and Adjustments
Force FX-8C Service Manual 5-3
Important
Activation tones sound to verify that the speaker is working
properly.
If the coag mode has been
optionally changed to default to
Desiccate or Spray, that
corresponding indicator illuminates
after the self-test is performed
successfully.
7.
If the self-test is successful, a tone sounds. Verify the following:
Indicators above the default mode buttons (Standard bipolar, Pure
cut, and Fulgurate coag) illuminate green.
Each display shows a power setting of one watt.
The REM Alarm indicator illuminates red.
If the self-test is not successful, an alarm tone sounds. A number may
momentarily appear in the Cut display and, in most cases, the
generator is disabled. Note the number and refer to Responding to
System Alarms in Section 6.