Using the RS-232 Serial Port
Setup, Tests, and Adjustments
Force FX-8C Service Manual 5-43
Commands for Displaying Data on a Computer
Command Result Displayed on Computer
:PP Front panel values
:PT Current real-time clock values
:PSC Date of last calibration
:PSA Total time activated in each mode
:PSB Number of times activated in each mode
:PSK Number of times output was activated
:PSP Average power setting used in each mode
:PSF List of system errors 1 through 10 with time and date stamp
:PSL List of system errors 11 through 20 with time and date stamp
:PSE# Number of times specified error has occurred
(# represents an alarm number listed in Section 6)
:PG Generator model and firmware version numbers