49
Verathon Service Partner Manual: Repair & Replacement
ATTACH NEW BATTERY EJECT BUTTON
5. Place the washer on the top cover.
Washer
6. Mount the filed push rod on the top cover.
Push rod
7. Place the rocker arm on top of the washer so that the knob of the push rod is located in the slot of the
rocker arm.
Rocker arm