28
PROCEdURE 8. CLEAN THE PRINTHEAd BY USING PAPER
In some instances when the printhead is producing irregular printouts, you can clean the printhead by
wetting a piece of paper with a cleaning agent and then running the paper through the printer. If this
procedure is successful, it is not necessary to disassemble the console in order to access the printhead
directly.
QUANTITY PART # DESCRIPTION
Parts
1 0800‑0042 Thermal paper roll for the printer
Tools
1 0400‑0039 BVI3000 battery
AR Isopropyl alcohol, 70% or higher
1‑3 Cotton swab
1. Place a charged battery into the console, and then press the Power button.
2. Open the paper roll door.
3. On the console, on the Main Configuration screen, press the RUN button. The console is now in
standard operating mode.
CABLE
TEST
PROG
PROBE
MFG
TEST
SETUP
NOW RUN
THIS UNIT IS NOT SETUP
4. On the Main Menu screen, press the PRINT button.
BVI 3000
SCAN ALT
VERATHON
PRINT NOTES
13:2705/22/14
5. On the Print screen, press the SEL button until TEST PRINT is selected.
PRINT SEL DONE
TEST PRINT
COST SAVING
LAST IMAGE
X

Table of Contents

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals