62
PROCEdURE 19. REPLACE THE CONSOLE LABEL
When reprinting the serial number, ensure that you reuse the original console serial number.
IMPORTANT
Complete this procedure if the console label is damaged or if you replaced the bottom cover.
When replacing a label, examine the existing console label and compare it to the following two figures in
order to determine whether or not the instrument is UDIcompliant.
In order to confirm that a unit is UDIcompliant, look for the following items of information on the label:
The unit’s date of manufacture
A DataMatrix code containing the unit’s ID and date of manufacture in machine‑readable form
When reprinting the serial number, ensure that you reuse the original console serial number.
Figure 20. UDIcompliant console label Figure 21. NonUDI console label
0570- 0090
012 3
0125-2530-00-40
Verathon Inc.
20001 North Creek Parkway
Bothell, WA 98011-8218 USA
(800) 331-2313 (USA & Canada)
+1 (425) 867-1348
Fax: +1 (425) 883-2896
Verathon Medical Europe B.V.
Willem Fenengastraat 13
1096 BL Amsterdam
The Netherlands
+31 (0) 20 210 30 91
Fax: +31 (0) 20 210 30 92
Medical - Ultrasound Equipment : ES60601-1:2005 and IEC 60 601-2-37:2007
MADE IN USA
0400-0039
0800-0042
B3123456
2016-05-31
(01) 0 0879123 00553 3
(21) B3123456
(11) 160531
Model: BVI 3000 0570-0090
0123
0125-0270
Verathon Inc.
Corporate Headquarters
20001 North Creek Parkway
Bothell, WA 98011-8218 USA
+1 800.331.2313 US & Canada
+1 425.867.1348
Fax: +1 425.883.2896
Verathon Medical
Europe B.V.
Linnaeusweg 11
3401 MS IJsselstein
The Netherlands
+31.30.68.70.570
Fax: +31.30.68.70.512
To locate your nearest distributor, visit us online at www.verathon.com. US Patent No. 4926871, 5235985 INTL Patents Pending
C US
Made in USA
B3123456
B3123456
QUANTITY PART # DESCRIPTION
Parts
AR 01250270 BVI3000 console label, non‑UDI
AR 0800‑0550 BVI3000 console label kit with serial number, UDIcompliant
Tools
1 Plastic scraper
All‑purpose cleaner
1 PC station with access to printer
1. If the instrument is UDIcompliant, order a UDI console serial number label kit (08000550).
If the instrument is not UDIcompliant, use a non‑UDI console label (0125‑0270).
2. If the bottom cover has an existing label, carefully peel it off.
3. Using a plastic scraper and an all‑purpose cleaner, remove any residue from the previous label.
4. If you are using a UDI console serial number label kit, skip to Step9
If you are using a non‑UDI console label, continue to the next step.

Table of Contents

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals