72
PROCEdURE 22. PROGRAM THE TAILCONE
QUANTITY PART # DESCRIPTION
Tools
AR 05700090 BVI3000 test console
*
1 0400‑0039 BVI3000 battery
*
See Part Numbers on page87 for orderable part number
1. Connect the fully assembled probe to a console that is operating in configuration mode. If needed, use
a test console and the procedure Enter Configuration Mode on page6 in order to put the system in
configuration mode.
2. Insert a charged battery, and then press the Power button on the console. The system turns on.
3. On the Main Configuration screen, press the PROG PROBE button.
CABLE
TEST
PROG
PROBE
MFG
TEST
SETUP
NOW RUN
THIS UNIT IS NOT SETUP
4. On the Program Probe screen, press PROG.
PRINTPROG SCAN
AA
5. On the probe, press the Scan button. This ensures that the correct probe type is set. Ensure the Phi
offset value is at 5.
6. Continue to the following procedure, Update the Probe ID.