EasyManuals Logo

Verathon BladderScan BVI 3000 Service Manual

Verathon BladderScan BVI 3000
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
61
Verathon Service Partner Manual: Repair & Replacement
INSTALL THE CONSOLE FEET
27. If not already present, apply double‑sided tape to the four areas for the console feet.
28. Place one foot tab into one of the two holes for the foot. If necessary, you may use a small instrument,
such as tweezers, in order to push the tab into the hole.
29. Use a small instrument to align the second prong, and then push the second prong into the other hole.
30. Ensure that the foot is securely attached to the console.
31. Repeat Step28 through Step30 in order to install all four feet on the console.
32. Do one of the following:
If you replaced the bottom cover or if the console label is damaged, continue to the procedure
Replace the Console Label on page62.
If you replaced the EPROM or PCB assembly, continue to Program the Console on page65.
If you do not need to replace the console label or program the console, continue to the chapter
Annual Certification on page6.

Table of Contents

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals