ATTACH THE BOTTOM COVER
22. Place the bottom cover on the overmold.
23. Using a Phillips bit, insert the (4) machine bottom cover screws (02610013) that secure the bottom cover
in place.
24. Ensure there are no gaps between the bottom cover, top cover, and handle. If needed, apply one or two
bar clamps to the sides of the console, and then confirm the gap is closed.
25. Using a Phillips bit, insert (2) self‑tapping bottom cover screws (02600283).
26. If used in Step24, remove the bar clamps.