84
12. If desired, repeat Step7 through Step9 two more times in order to ensure measurement consistency.
13. When you have collected sufficient scans in order to confirm aiming accuracy, press the DONE button.
DONE SCAN X
READY TO SCAN
132ml
CURRENT = 132ml
CONFIRM MEASUREMENT ACCURACY
14. Confirm that the measurement results are within the accuracy range for your phantom as defined in
Accuracy Specifications on page85.
15. If the measurement(s) are consistent and accurate, then the instrument is performing as intended.
If the instrument is unable to produce consistent, accurate measurements, service is required on the
instrument. Continue to the Diagnostics & Troubleshooting chapter on page18.