EasyManua.ls Logo

Vertiv Tech SmartCabinet2-E Series - Page 48

Default Icon
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SmartCabinet2-E Series Computer Room Solution Product User Manual
42 Chapter 3 Operation Settings
1. Menu Item
2. Controlled state
3. Each level the number of real-time alerts
4. System Title
5. Logo
6. Logout [current user]
7. Real-time alarm display list
8. Alarm alert settings
9. Time Calibrating Link
Figure 3-35 Intelligent Monitoring System Homepage
Notices
When the system is abnormally running, a corresponding fault alarm will be generated. The emergency ventilation system can
prevent the temperature inside the cabinet from rising continuously in a short time. If the high temperature fault alarm does not
cease for a long time, the user needs to open the front and rear doors of the cabinet on the site, and contact the service personnel
to check the fault.
Click the Setting button in the Figure 3-36 to enter the setup status of the home page, as shown in the Figure 3-36.
Figure 3-36 Set status
After the home page enters the setup state, the setup method is as follows:
1) Model setting
Click the Select Type button to pop up the page as shown in the Figure.3-37.
Click the Left(<) and Right(>) buttons to select the model image.
After selecting, click the Select Type button and then click “OK”, the homepage will display the background image.

Table of Contents