Summary of Error Messages 87
5.1.2 Warning Messages
Overview
Warning messages appear on the 600 controller screen to instruct you to:
• Enter acceptable values
• Check your entries
Pay attention to warning messages and take corrective action. Failure to do so may cause
the 600 controller to produce erroneous results or to fail to perform your chromatographic
run.
Resuming Programming or Operation
To resume programming or operation after a warning message appears, you can either:
• Press Clear.
• Wait approximately 5 seconds until the message disappears.
Table 5-1 lists the warning messages shown on the 600 controller screen.
Table 5-1 Warning Messages
Message Displayed Message Appears Meaning
An Inject
Occurred
During operation in the
Operate Gradient or
Operate Method screen
Informational message indicating
that an inject signal has been sent.
Cannot Save Table
Currently
Executing
While running from the
Operate Method screen
The table number you intended to
save for a new or modified table is
the same as the table in process.
Abort the run or wait until finished
to replace the table with the modi-
fied table.
Check Corre-
sponding Gradient
Table/Events
Table
While programming a
Gradient or Event table
Because the Gradient and Event
tables always run together (dictated
by the table number), any modifica-
tions made to the Gradient table
should be consistent with the Event
table of the same table number.