2 Unpacking and Preparing Your System
Reporting Damage
Immediately report any damage to both the shipping carrier and to your Technical Service
Representative. You can contact Waters Technical Service at 800 252-4752, Canadian and
U.S. customers only. Other customers, call your local Waters subsidiary or call Waters
corporate headquarters for assistance in Milford, Massachusetts (U.S.A.).
If any items are damaged, use the shipping container for subsequent claim purposes.
1.2 Selecting the Site Location
Site Selection Requirements
Install the 600E Multisolvent Delivery System in an area where:
Temperature is 4 to 38 °C (39 to 100 °F). Avoid placing in direct sunlight, or near
heat registers or air-conditioning vents.
Relative humidity is 20 to 90% non condensing.
Available bench space for the 600E system includes:
Waters 600E Pump Physical Specifications
Waters 600E Controller Physical Specifications
Parameter Specification
Height 9.0 in. (22.86 cm)
Length 22.0 in. (55.88 cm)
Width 11.0 in. (27.94 cm)
Weight 47.5 lbs. (21.56 Kgm)
Parameter Specification
Height 7.0 in. (17.78 cm)
Length 21.25 in. (53.97 cm)
Width 11.25 in. (28.57 cm)
Weight 13.0 lbs. (5.90 Kgm)