EasyManua.ls Logo

Waters 600e - Making Fluidic Connections to an Autosampler

Default Icon
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Making Fluidic Connections to an Autosampler 25
Figure 2-15 RCM 8 x 10 Orientation
4. Pass the cartridge outlet tubing through the notch in the pump drawer front panel.
Use the shortest possible length of 0.009-inch ID tubing possible from the column
outlet to the detector cell inlet.
5. Pump eluent through the RCM 8 x 10. Verify that there are no leaks at the inlet or
outlet connections.
For detailed operation information, refer to your Waters RCM 8 x 10 Operators Manual.
2.4 Making Fluidic Connections to an Autosampler
You have the option to use the Waters 600E Multisolvent Delivery System with an
external autosampler. You can:
Connect an autosampler in series with the Rheodyne 7725i manual injector
Connect an autosampler in place of the Rheodyne 7725i manual injector
Note: When using a non-IEEE-488 autosampler, you must also attach a cable from
the autosampler to the Inject terminal (on the 600 controller) to instruct the 600
controller to acquire data or run methods and tables. See Section 3.5.2, Connecting
a Non-IEEE-488 Autosampler.

Table of Contents

Other manuals for Waters 600e

Related product manuals