Summary of Error Messages 91
5.1.3 Shutdown Messages
Overview
Shutdown messages appear on the 600 controller screen when the 600E pump:
Exceeds an operator-set limit
Receives a Stop Flow signal from an external device
All shutdown messages require immediate user action (see Table 5-2).
Unable To Setup
Injector
During operation in the
Direct Control or
Operate Methods
screen
The 600 controller is unable to set
up the autosampler due to one of the
following:
Autosampler is not turned on
Carriage door is open
Sample carriage is not locked in
place
Autocalibration has not been
performed
Resolve the problem, then go to the
System Configuration screen and
rescan the IEEE-488 bus.
UV Detector 1 Not
Found
UV Detector 2 Not
Found
During operation in the
Direct Control or
Operate Methods
screen
The UV detector (1 or 2) has disap-
peared from the IEEE-488 bus.
Verify if the UV detector:
•Is turned on
Was turned on before the 600
controller
Go to the System Configuration
screen and rescan the IEEE-488
bus.
Value Entered Is
Out Of Range
While entering param-
eter values in a screen
The entered parameter value is out
of the allowable range. Check the
range for the parameter. Press the
Help key for a summary of allow-
able ranges.
Table 5-1 Warning Messages (Continued)
Message Displayed Message Appears Meaning