90 Error Messages, Diagnostics, and Test Procedures
RI Detector Not
Found
During operation in the
Direct Control or
Operate Methods
screen
The 410 detector has disappeared
from the IEEE-488 bus. Verify if
the 410 detector:
•Is turned on
• Was turned on before the 600
controller
Go to the System Configuration
screen and rescan the IEEE-488
bus.
Step Does Not
Exist
While running from the
Operate Method screen
A step number that was inserted
into the initial Operate Methods
screen does not exist in the Program
Method screen.
Return to the Program Methods
screen and verify the program
methods table.
Table Being
Stored...
Please Wait
While saving a
Gradient table, Event
table, or Detector table
Keyboard is locked while the Save
function is storing the new table.
Table Is Full While programming the
program methods table
Exceeded the maximum 48 steps
for this table. Determine which
lines need to be deleted.
Unable To Set Up
Detector
During operation in the
Direct Control or
Operate Methods
screen
The detector has not been turned on
and calibrated. Turn on detector and
wait for calibration to complete.
(This also applies if the lamp is
turned off.)
Go to the System Configuration
screen and rescan the IEEE-488
bus.
Table 5-1 Warning Messages (Continued)
Message Displayed Message Appears Meaning