Summary of Error Messages 89
Invalid Injec-
tion Number
While running from the
Operate Method screen
The Injection Number is not part of
the Step indicated in the Operate
Method screen.
Return to the Program Methods
screen and verify the program
methods table.
Invalid Vial
Number
While running from the
Operate Method screen
A vial number was selected that is
not included in the selected step of
the Step indicated in the Operate
Method screen.
Return to the Program Methods
screen and verify the program
methods table.
Line is Incor-
rect Or
Incomplete
While programming a
Gradient table, Event
table, Detector table, or
the Program Methods
table
You did not enter all required
entries for a specific line. Enter all
required entries before attempting
to move the cursor to a new
location.
Number Out Of
Range
While entering param-
eter values in a screen
Entered value is out of the allow-
able range. Check the range for the
parameter. Press the Help key for a
summary of allowable ranges.
Replace with New
Table
1 = Yes 0 = No
While saving a
Gradient table, Event
table, or Detector table
Abort the save operation if the table
associated with the number entered
should not be erased.
Rear Panel Stop
Flow Input
Detected
After 600 controller
receives a signal at the
Stop Flow terminal
Suspend 600E system operation
until the source of the Stop Flow
signal is removed.
Table 5-1 Warning Messages (Continued)
Message Displayed Message Appears Meaning