Making Fluidic Connections to a Detector 27
2. Disconnect the Rheodyne 7725i manual injector outlet from the column or column
heater inlet in the pump slide-out drawer (see Section 2.3, Installing a Column or
Cartridge).
3. Connect the autosampler outlet line to the column or column heater inlet in the
pump slide-out drawer (see Section 2.3, Installing a Column or Cartridge).
4. Disconnect the Rheodyne 7725i manual injector inlet from the transducer outlet
line. Connect the autosampler inlet line with a union to the transducer outlet line.
2.5 Making Fluidic Connections to a Detector
To connect the 600E pump to your detector inlet:
1. Pass the column or cartridge outlet tubing through the notch in the pump drawer
front panel. Use the shortest possible length of 0.009-inch ID tubing from the
column outlet to the detector cell inlet. For details on attaching tubing to a column
or cartridge, refer to Section 2.3, Installing a Column or Cartridge.
2. Pass the waste line from the detector cell outlet to a proper receptacle.
For details on making fluidic connections to your detector, refer to the associated detector
operators manual.