Making IEEE-488 Interface Connections 39
Note: You cannot run methods or method sets, or use the QuickSet Control window
unless all devices on the IEEE-488 bus that are assigned to a system on the
Millennium Chromatography Manager are powered up and calibrated. You do not
need to power up instruments that are not assigned to a system, or that are
assigned to a system you do not intend to use.
4. Power up the printer and monitor.
5. Power up the computer.
PowerLine Controller Powerup Sequence
Perform the following powerup sequence when using the 600 controller as a PowerLine
Controller:
1. Power up all equipment not controlled through the IEEE-488 interface.
2. Power up all equipment controlled through the IEEE-488 bus.
3. Power up the 600 controller.
Power Down Sequence for IEEE-488 Devices
If you do not plan to use the Waters 600E system for a long period of time (overnight or
longer), power down the system.
Note: For a list of power-down precautions for storing the 600E pump, refer to the Waters
600E Multisolvent Delivery System Users Guide, Chapter 3, Preparing Your 600E
Multisolvent Delivery System for Operation.
To power down the 600E system:
1. Ensure the system is purged of salts.
2. Stop your chromatographic run (if necessary).
3. Press the On/Off switch (on the 600 controller front panel) to power down the 600E
system.
4. Power down the other IEEE-488 devices.