EasyManua.ls Logo

Waters ACQUITY UPC2 - Page 108

Default Icon
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
November 16, 2016, 715004521 Rev. C
Page 108
See also: For detailed instructions, see the console online Help.
Recommendation: Reboot the workstation after configuring software for a column
manager.
Result: After completing this procedure, when you select each column manager in the console
system tree, the columns appear. The primary column manager is listed first with Columns 1
through 7 available. The secondary column manager is listed second with Columns 1 through 8
available.
Exception: To run the ABPR health test, select Bypass for both column managers.
See also: “Running an ABPR health test when the system includes a CM-30S”on page 77.
C.3.0.3 Select a column
You can use only one column at a time. The software is programmed to limit your selection to one
column. An alarm will appear if you select one column in each column manager.
You select a column in the instrument method editor or in the control panel. Then, for the column
manager that does not control this column, you select None. This stops the flow of sample and
solvent to the unused column manager.
Example: Select a column in the primary column manager, then select None in the secondary
primary manager.
C.3.0.3.1 Selecting a column before running system startup
Requirement: Ensure that you select a valid column combination before you run the system
startup feature.
To select a column before running the system startup:
1. From the Console tree, select the system.
2. From the Control menu, select Start up system.
3. On the Equilibrate to Method tab of the System Startup dialog box select the CM and CM-2
subtabs.
4. For the Column Selection, select the column number or select None.
5. Make any needed changes on the other subtabs, and then click Start to begin the system
startup.
C.3.0.3.2 Selecting a column in instrument method editors
To select a column in instrument method editors:
1. Select the column manager instrument method editor that controls the column.
Tip: In most cases, in the instrument method editor, the primary column manager is named
ACQ-CM and the secondary column manager is named ACQ-CM-2.
2. Select the General tab.
3. Select the Column Selection mode.
4. In the Column Selection area, in the Valve Position field, select the column number.

Table of Contents

Other manuals for Waters ACQUITY UPC2

Related product manuals