November 16, 2016, 715004521 Rev. C
Page 59
Table 5–3: Core 1 configuration with a PDA, TQ-XS, and splitter
Tube Tube description From To
ID
Notes:
recommended
trim length
1 750 mm (29.5 inch)
tool-free ESI (or
APCI) probe
capillary
(connects to
external, V-detail
fitting on splitter)
(Note: To connect
the tool-free ESI (or
APCI) probe
capillary to the MS
splitter module,
remove the
29.5 inch PEEKsil
tubing that is
pre-connected to
the module’s
external V-detail
fitting and connect
the probe capillary
in its place.)
TQ-XS inlet
probe
Splitter module 50 µm Do not trim
2 50 inch
PEEK tubing
(pre-connected to
the splitter)
Splitter
module
Left side of the
ccManager tee
0.010 inches Do not trim more
than 23 inches
359inch
PEEK tubing
(pre-connected to
the splitter)
Top of the
PEEK union
on the 515
makeup pump
Splitter module 0.005 inches Trim as required
4 26 inch
PEEK tubing
(pre-connected to
the splitter)
CM-A or PDA Splitter module 0.007 inches Do not trim