November 16, 2016, 715004521 Rev. C
Page 64
Table 5–7: Core 2 configuration with a PDA, SQD, and splitter
Tube Tube description From To
ID
Notes:
recommended
trim length
1 29.5 inch PEEKsil
tubing
(pre-connected to
the splitter)
Splitter module Mass spectro-
meter
50 µm Do not trim
250 inch
PEEK tubing
(pre-connected to
the splitter)
Splitter module Left side of the
ccManager tee
0.010 inches Do not trim more
than 23 inches
359inch
PEEK tubing
(pre-connected to
the splitter)
Top of the
PEEK union
on the 515
makeup pump
Splitter module 0.005 inches Trim as required
446inch
PEEK tubing
(pre-connected to
the splitter)
The CM-A or
PDA
Splitter module 0.007 inches Do not trim more
than 20 inches

Table of Contents