TROUBLESHOOTING GUIDE
In the event of an error code, unplug the 3M™ Attest™ Autoreader 490 and plug it back in to initiate the selfdiagnostic test. If error code reoccurs, call for repair or replacement.
The LCD Display panel will indicate detected malfunctions by displaying a caution or error code under affected BI wells. The audible alarm, if active, sounds and the LCD Display
backlight flashes to alert the user to all caution (except C1) and error codes. The alarm and backlight can be silenced and extinguished, respectively, by pressing the Buzzer Alarm OFF
button
. Note that disabling the audible alarm does not disable the flashing LCD Display backlight. Refer to the following:
Code Problem Solution
C1
Unit Warmup Caution
Incubation block not at temperature.
Allow a 30 minute warmup after 3M™ Attest™ Autoreader 490 is plugged in. (Note: 3M™ Attest™
BI readings are disabled until incubator block reaches 60ºC +/2ºC).
C2 BI Removed from Well Caution
3M™ Attest™ BI has been removed before incubation
iscomplete.
Replace 3M™ Attest™ BI in proper well within 10 seconds to prevent loss of collected data.
C4 Ambient Light Caution
Move 3M™ Attest™ Autoreader490 to reduced ambient lightconditions.
E1 Temperature Control Error
This error will occur if system is no longer able to control the
temperature of the incubator block (60ºC +/2ºC).
Unplug the unit and plug it back in to clear error.
If error doesnt clear, call for repair or replacement.
E3 Memory Error
3M™ Attest™ Autoreader490 unable to save/read data
frommemory.
Unplug the unit and plug it back in to clear error.
If error doesnt clear, call for repair or replacement.
E7 Communications Error
This error will occur if internal communications havefailed.
Unplug the unit and plug it back in to clear error.
If error doesnt clear, call for repair or replacement.
E8 LED Error
This error occurs for a specific well if the system detects that a
UV LED or Optics Detection circuit is notworking.
Unplug the unit and plug it back in to clear error.
If error doesnt clear, call for repair or replacement. Do not use affected incubation well.
E11 BI Inserted Incorrectly Error
Incubation time remaining will not display and fluorescent
readings are nottaken.
Remove BI and inspect to verify cap is fully closed, glass ampoule is crushed, and growth media is in
growth chamber. If necessary, repeat BI activation steps and return BI to incubation well, ensuring it is
seated properly.
E12 Power Failure Error
Power to Autoreader was off for too long to provide valid result
for affected BI(s).
Ensure power supply connection to Autoreader is secure. Retest sterilizer with a new BI.
E14 Ambient Light Error (occupied well)
Fluorescent result will notdisplay.
Rely on optional visual pH color change result for BI in affected well. Move 3M™ Attest™ Autoreader
490 to reduced ambient light conditions to avoid future occurrences. To clear E14 error code, after
visual pH color change result has been documented and BI has been removed, unplug the 3M™
Attest™ Auto-reader 490 and plug it back in.
Other Unwanted alarm sounds when positive BI detected.
Disable alarm, refer to 3M Attest Autoreader 490 Configuration section for instructions.
Other Alarm Sounds
3M™ Attest™ Biological Indicator has been removed from
incubation/reader well prematurely.
Replace 3M™ Attest™ Biological Indicator in incubation/reader well within 10 seconds to prevent
loss of data collected.
Other LCD Panel does not display properly. Cycle power. If problem persists call for repair or replacement.
Other Negative Control BI
BI not properly activated. Media did not contact spore carrier.
Control BI was sterilized.
Ensure BI cap is closed and glass ampoule is crushed.
Flick BI until media flows to spore chamber at bottom of BI.
Use a nonsterilized BI as a control.
Check process indicator on BI cap.
Other Unexplained Positive BI
Adhesive residue on vial.
Test vial is absorbing fluorescent residue from a chemical
indicator or tape.
Do not place tape or labels on vial.
Place the vial so it does not come in direct contact with chemical indicators or tape.
Other
3M™ Attest™ Autoreader 490 will not enter Configuration
mode. Pushbutton Configuration may be disabled on 3M™
Attest™ Autoreader 490 Web App.
On 3M™ Attest™ Autoreader 490 Web App Configuration Page, go to Pushbutton Configuration and
select the Enable radio button. Save changes to begin using new setting.
Other
3M™ Attest™ Autoreader 490 LCD display flashes and
asterisks (***) and/or E11 and/or C4 codes appear on display
below unoccupied incubation wells.
Move 3M™ Attest™ Autoreader 490 away from sunlight or sources of incandescent light.