SPECIFICATIONS
Power Supply Specifications
Power supply Operating Condition Units
Voltage Range 100240 Volts AC
Frequency 50/60 Hertz
Current 0.6 Amps
Environmental Operating Conditions
Environmental Condition Operating Condition Units
Altitude 3000 (max) Meters
Operating Temperature 16 40 ºC
Storage Temperature 10 to 50 ºC
Operating Relative Humidity 2080 (noncondensing) %
Voltage Range 12 Volts DC
Frequency DC
Current 1.2 Amps
Installation/Over Voltage Category II
Pollution Degree 2
INSTRUCTIONS FOR USE Initial Setup
ChooseLocation
Place 3M™ Attest™ Autoreader 490 on a firm level surface away from sunlight and
incandescent light. To save counter space, units may be stacked two high.
Plug In Unit
Connect the power cord to the power supply. Connect the power supply to the 3M™
Attest™ Auto-reader 490 and plug the power cord into an AC grounded outlet.
WARNING:
•
Do not place the instrument in environment exposed to sunlight or strong
incandescentlight.
•
Do not place instrument close to any device that emits a strong
electromagneticfield.
•
Do not use on a vibrating surface.
Wait 30Minutes BeforeUsing
Once plugged in, allow a 30 minute warmup period before placing biological indicators
into the incubation wells. The C1 caution code will disappear when proper incubation
temperature is reached.
SpecialConsiderations
It is recommended that the unit be left on to eliminate warm-up periods. The unit
may be turned off when it is anticipated that it will not be used for an extended period.
WARNING: Use only the power supply specified for this product and certified for
the country of use.