EasyManua.ls Logo

ABB FSM4000 - Page 68

ABB FSM4000
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
68 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
8 Operation
Parameter descriptions
Submenu / parameter Value range Remark
Switching threshold
Numerical entry
Menu visible when ‘New adjustment‘ is selected in the ‘DLR Mode
parameter. Switching threshold for turning off the ‘Empty pipe’
function. The switching threshold is calculated by the transmitter
from both values ‘Adj. DLR empty’ and ‘Adj. DLR full”.
The calculated value can be changed here for fine adjustment in
the range from 100 to 1000000.
Not displayed, if DEP is switched off or DEP mode is standard.
100000
Alarm e. pipe:
Off
If the alarm is switched on, the empty pipe contact is actuated
when the pipe empties. The ‘empty pipe‘ or ‘General alarm’
switch output must be selected in the ‘Switch contacts’
submenu. When the function is turned off, the alarm does not
respond.
Not displayed, if DEP is switched off.
Off A
Iout for e. pipe
Low Alarm
High alarm
0 %
Set the current output value to Low or High in event of error.
Not displayed, if DEP is switched off.
Low Alarm

Table of Contents

Other manuals for ABB FSM4000

Related product manuals