86 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
8 Operation
Parameter descriptions
Submenu / parameter Value range Remark
Submenu
Set the time-out limits and alarm behavior for values over or
under the coil resistance.
Notice: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see ‘Submenu
Adjustment’ on page 93.
Alarm Coil Resist
Min. value
0 ... 1500
Lower range limit for alarm.
2 ohms
Max. value
0 ... 500
Upper range limit for alarm.
500 ohms
Message
Off, warning, error
Defines the alarm behavior.
Off Notice: If coil resistance is being measured (at least every 60
s), the most recently acquired flow rate is retained during the
measurement (2 s approx.).