FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 91
Submenu / parameter
Value range
Remark
Submenu Sets the status of all alarms to ‘Off’ and the alarms to the
default values.
Alarm Factory Set
Accept Yes, exit via C/CE
Yes → Enter
Submenu
Wired-on. Failure
Submenu
Defines alarm behavior in case of an accidental coil ground.
CoilShtCircGnd
Notice: If the coil circuit is being checked (at least every 10
minutes) for an accidental ground, the most recently acquired
flow rate is maintained (approx. 2 s).
Message
Off, warning, error
Off
Submenu Defines alarm behavior in case of a short circuit or a wire
break in the electrode circuit.
Notice: This function can only be used after the electrode zero
has been adjusted. For additional information, see ‘Submenu
Adjustment’ on page 93.
Electrode State
Message
Off, warning, error
Off