EasyManua.ls Logo

Abbott AEROSET - Absorbance Mode

Abbott AEROSET
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution Assays/Reagents
AEROSET
®
System Troubleshooting Guide 29
94816-107—November 2004
Absorbance Mode
When the system is in Absorbance mode, results print as
absorbance values. A result equal to or higher than the
Multi DOA Cutoff calibrator absorbance is considered positive
for the assay tested and negative for a value lower than the
Multi DOA Cutoff calibrator absorbance.
Perform the following steps to configure an assay in
Absorbance mode.
1. Select <Assays> in the Information Access Area of the
Main Display. The ASSAYS screen displays.
2. Select the desired assay button. The ASSAY STATUS
screen displays.
3. Select <Config> in the right column of the screen. The
Assay Configuration screen displays.
4. The Outline page is the default view.
Figure 5: Assay Configuration Screen, Outline Page
5. Verify Quantitative Ranges is selected (). This is the
default setting.
6. Edit the L-Reference field from “-9999” to “0.0”.
Select

Table of Contents

Related product manuals